Monitoramento da taxa de saturação de oxigênio via método de magnificação de vídeo euleriana

Peripheral oxygen saturation (SpO2) monitoring is critical during surgeries and in intensive care units. Over the last decades, several studies have addressed noncontact monitoring of vital signs, but few reports have focused on SpO2. This study presents an SpO2 monitoring method without physical co...

ver descrição completa

Autor principal: Rosa, Alessandra de Fátima Galvão
Formato: Dissertação
Idioma: Português
Publicado em: Universidade Tecnológica Federal do Paraná 2019
Assuntos:
Acesso em linha: http://repositorio.utfpr.edu.br/jspui/handle/1/4320
Tags: Adicionar Tag
Sem tags, seja o primeiro a adicionar uma tag!
Resumo: Peripheral oxygen saturation (SpO2) monitoring is critical during surgeries and in intensive care units. Over the last decades, several studies have addressed noncontact monitoring of vital signs, but few reports have focused on SpO2. This study presents an SpO2 monitoring method without physical contact with the patient using imaging photoplethysmography. Eulerian video magnification - (EVM) was employed in the proposed solution to amplify the changes in skin color due to the heart cycle. The developed software operates on a widely available, lowcost hardware platform. Experiments were performed with nine healthy individuals. When comparing the experimental results to those obtained with a reference pulse oximeter, the proposed solution showed good accuracy in SpO2 measurements, with a difference of less than 2%.